Multi-purpose operating table
Inquiry Basket > > > > Inquiry Now
  • Product Name:

    Multi-purpose operating table

  • Model NO.: RT50202

Description:
Multi-Purpose Operating Table
   This table are compactly designed and sturdily constructed, and may be raised or lowered hydraulically by means of an oil pump. They are suitable for surgical procedures of the head, neck, thorax, abdomen, perineum and extremities as well as for gynaecological, obstetrical, urological and oto-rhino-laryngological operations. Various positions can be quickly and safely adjusted.
  The controlling devices, hand wheels and crank handles are mounted at the head-end of the table. All positions of the table are controlled entirely from the head-end.
  Chassis inside  & outside cover of model 3008B are stainless steel .The structures and technical parameters are the same.
1. Specifications:
Dimensions of Table Top :  
    Overall length ………………………………………………………2040±40(mm)
    Width…………………………………………………………………480±20(mm)
Dimensions of Sections :
    Head Rest……………………………………………………………280×320(mm)
    Back section…………………………………………………………500×480(mm)
    Kidney Bridge…………………………………………………………84×470(mm)                     
    Seat Section………………………………………………………… 370×480(mm)                        
    Leg Section……………………………………………………………500×480(mm)                 
    Foot Rest………………………………………………………………200×535(mm)
Table Top from Floor :
    Max. Height…………………………………………………………1040±50(mm)
    Min. height……………………………………………………………800±50(mm)
Range of Movements :
    Max. Reversed Trendelenburg……………………………………………………≥50°
    Max. Trendelenburg………………………………………………………………≥30°
    Max. Lateral Tilt …………………………………………………………………≥20°
Adjustments of  Sections :
    Head Rest-raised from the horizontal…………………………………………… ≥45°
    Head Rest-lowered from the horizontal…………………………………………≥90°
    Back Section-raised from the horizontal………………………………………… ≥80°
    Back Section-lowered from the horizontal…………………………………………≥8°
    Kidney Bridge Elevation………………………………………………………≥120mm
    Leg Section-lowered from the horizontal……………………………………………≥90°
Foot Rest-can be kept horizontally or perpendicularly to the Leg Section.
2.Controlling Devices of Head-End Controlled Table:
   1、Trendelenburg position, reversed Trendelenburg position and tilt in left or right direction are regulated by one single Hand wheel (2). The movements of the back section and leg section are controlled by Hand Wheel (3). For directions for use please see the label attached to the crank sleeve.
 2、The kidney bridge is controlled by Crank Handle (1) situated at the center
 .3、Both the ascending and descending movements of the table top are controlled by the oil pump. When the products leave the factory, the oil pump has been pumped into the pail, so
 the oil pump can not ascend. The Consumers must move the cover of the filling aperture
  at the fop of the oil pump. Use a funnel to refill the oil pump slowly. By pressing down Pedal (4) repeatedly, Press once, the table top will rise up a certain height. When the table top be elevated to the highest position, it can be stops automatically,when the same pedal is pressed to the dead end and not loosen,the table top will begin to descend slowly and gently.
   4、The moving and stabilization of the base. By loosening Handle (5) of the base clockwise, the table can move. On the contrary, it can be stabilized.



3. Accessories:
         Accessories     Pc(S)
  Anesthetic Screen    1
  Shoulder Rest    2
  Arm support    2
  Body Support, Large    1
  Body Support, Small    1
  Knee Crutch    2
  Cushion    1
  Body Restraint Strap    1
  Wristlet    2
4. Cautions:
1、The Head-rest adjustment is adopted by a strict teeth lock. If the position of the head-rest needs adjusting, the teeth lock pole must be first separated and adjusted to suitable position, the got loose, otherwise it is easily damaged.
  2、Lubricating oil should be added to every active part, so that the bed can be operated easily and prolong their life. Smear Vaseline over the tracks of the accessories on each side of the bed and the racks under the table, thus preventing it from being rusty.
  3、Before putting into use, all attachments should be checked to see if they are locked securely so as to prevent them from loosing or sliding off while used.
  4、As the bed has many parts, keep all the accessories in good order.
  5、After operating every time, clean the dirt on the bed, keep the bed and accessories clean and prolong their life.
  6、The ground must be kept dry. While the floor is washed and scrubed, please try not to make water spill into the plate or on the rotated wheels, thus preventing it from being rusty.
  7、It must be repaired and kept clean regularly once a year.
  8、In the clinical use, the lateral lean angle of the operating table is bigger than 8 degrees (the standard is equal or bigger than 20 degrees), the user should think the possibility that the patient falls from the table top, and selects a body frame and body belt and so on to ensure the use safety.
  This device in package should be stored indoor with a relative humidity less than 80%, no erosive air and good ventilation is required. No side-sloping.
It can be shipped in regular transportation without crash, oscillation, rain or direct sunlight.

Feedback Form
*Email
*Content
*Product Name
*Your Name
*Company
*Tel./Fax.

Related Products:

Address:Moon Lake Tower, NO.16, Nanzhan East Road, NnigBo, CHINA 中国
Phone:0086-574-27899000, 27899002  Fax:0086-574-27899001  E-mail: Hot Products
CopyRight © 2010-2012 NingBo Resto Medical Instruments Co.,LTD All Rights Reserved.